• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE FB CR TRL SZ 5 12MM; TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE FB CR TRL SZ 5 12MM; TIBIAL TRIALS Back to Search Results
Model Number 2545-05-220
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2023
Event Type  malfunction  
Event Description
It was reported that the cement in groove cannot be removed.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary the device associated with this report was returned for analysis.Visual analysis revealed that one of the tabs was found chipped.No foreign substances were observed, therefore the report allegation cannot be confirmed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE FB CR TRL SZ 5 12MM
Type of Device
TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key16589387
MDR Text Key311740543
Report Number1818910-2023-06313
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295479444
UDI-Public10603295479444
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/22/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model Number2545-05-220
Device Catalogue Number254505220
Device Lot NumberBFA239V
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/09/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/03/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-