Model Number HL-90 |
Device Problem
Leak/Splash (1354)
|
Patient Problems
Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/27/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: no information has been provided to date.Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when required.
|
|
Event Description
|
It was reported that the device was leaking from the bottom of the reservoir.There has been no report of patient involvement or no observable clinical symptoms or a change in symptoms identified in the patient.
|
|
Manufacturer Narrative
|
Other, other text: b5 and h6 codes updated.Device evaluation: one device was returned for investigation in used conditions.Visual inspection noted a cracked water tank.Functional testing was done where the customer complaint was confirmed.The device was leaking from the reservoir due to a crack in the tank cover.The gasket and tank cover were replaced.The device was tested again and passed all functional tests.A device history review was not done as the device is beyond a year from the manufacturing date.Service history review identified that this device had not been in for service previously.
|
|
Event Description
|
Additional information: the event occurred during testing.No patient was involved.
|
|