MAUDE Adverse Event Report: DEPUY SPINE INC CONFIDENCE HALF DOSE KIT (5CC); CEMENT, BONE, VERTEBROPLASTY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SPINE INC CONFIDENCE HALF DOSE KIT (5CC); CEMENT, BONE, VERTEBROPLASTY Back to Search Results
Model Number 283905000
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/28/2023
Event Type  malfunction  
Event Description
It was reported that on february 28, 2023, upon opening a large confidence cement kit and following the precise mixing instructions, the cement seemed to harden and when attaching the vial that is supposed to fill with cement, the vial did not fill and the cement appeared still grainy instead of smooth.A second large confident from the same lot.Again, the same issue occurred where the vial did not fill and the cement looked incorrect.All ingredients were mixed as measured and timed when mixing.Lastly, we opened a half dose 5cc confidence kit, and the goal of liquid was broken and unusable.No contact was made with patient.The surgery was surgery was successfully completed with surgical delay of 20 minutes.This complain involves three (3) devices.This report is for one (1) confidence half dose kit (5cc) this is report 2 of 3 for complaint (b)(4).
 
Manufacturer Narrative
Product complaint #(b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply.Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a j&j sales representative.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h4: part # 283905000 lot # 326233 supplier: affiliate selzach - batch1: lot qty of 02 units were released on (b)(6) 2022 with no discrepancies.- batch2: lot qty of 03 units were released on (b)(6) 2022 with no discrepancies.- batch3: lot qty of 05 units were released on (b)(6) 2022 with no discrepancies.- batch4: lot qty of 08 units were released on (b)(6) 2022 with no discrepancies.- batch5: lot qty of 01 units were released on (b)(6) 2022 with no discrepancies.- batch6: lot qty of 19 units were released on (b)(6) 2022 with no discrepancies.- batch7: lot qty of 07 units were released on (b)(6) 2022 with no discrepancies.No ncrs were generated during production.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIDENCE HALF DOSE KIT (5CC)
Type of Device
CEMENT, BONE, VERTEBROPLASTY
Manufacturer (Section D)
DEPUY SPINE INC
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key16592977
MDR Text Key312224857
Report Number1526439-2023-00564
Device Sequence Number1
Product Code NDN
UDI-Device Identifier10705034209555
UDI-Public(01)10705034209555
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2023
Device Model Number283905000
Device Catalogue Number283905000
Device Lot Number326233
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2023
Initial Date FDA Received03/22/2023
Supplement Dates Manufacturer Received04/28/2023
05/03/2023
Supplement Dates FDA Received05/02/2023
05/18/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/10/2022
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
CONFIDENCE KIT, NO NEEDLES; CONFIDENCE KIT, NO NEEDLES
-
-