It was reported that on (b)(6) 2023, a 29mm sjm masters series valve expanded cuff was chosen for implant.While attempting to implant the valve, the leaflet of the valve was broken.A new 29mm epic valve was then successfully implanted as a replacement.The patient was reported as stable.
|
An event of a dislodged leaflet was confirmed.The device was returned to abbott for investigation, and one leaflet was dislodged and fractured while the other leaflet was properly inserted into the orifice.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and that the product met all defined manufacturing specifications.How or when the damage occurred could not be conclusively determined.However, all fractured pieces were returned, and the damage may have been caused by some external force applied to the valve, which overstressed the carbon material.
|