The apc/esu system with it footswitch were returned and thoroughly inspected/tested.The findings were as follows: apc: a technical safety check was performed on the unit.This included an electrical safety check, a functional check of the equipment's features, and a power output check.All features were/are functioning properly within specifications.Also, the gas flow rates were measured and found to be within their acceptable ranges.In addition, no anomalies were found in the device history record (dhr).In conclusion, no equipment problem was found that would have caused or attributed to the event (note: unrelated to the reported issue, the unit's software was updated.).Esu: no issues were found with the unit that would have caused or contributed to the reported patient incident.The evaluation included an electrical safety check, a function check of each of the equipment's features, and a power output check.All features were/are functioning properly within specifications.In addition, no anomalies were found in the device history record (dhr) of the involved device.However, the esu needed servicing (i.E., a monopolar socket that is not involved with apc was worn and numerous error codes were found in the unit's error log over the last year); but it was not serviced due to its age (note: the model software version is obsolete.).Vio double foot pedal with remode (part number 20189-105 and lot number zv-zv) the accessory was found to be functioning properly.However, its cable was worn and needed to be replaced.Most likely, there were many factors involved in the reported incident.The desired setting may not haven selected/used, etc.However, the patient being elderly, a very thin-walled area needed to be treated, etc.Were key patient factors in the event.Nevertheless, upon the interventional work, the remaining tissue of the bowel did not stay intact which resulted in the perforation.Finally, no conclusive determination could be made as to the cause of the incident.To further address the issue, additional in-service work will be offered to the staff at the medical center.No trends have been identified.Erbe usa, inc.Is now closing the file on this event.
|
It was reported that an erbe system, argon plasma coagulator (apc)/electrosurgical unit (esu) [model vio 300 d, part number (p/n) 10140-000, serial number (b)(4)] system was involved in a patient incident.The apc/esu system with an apc circumferential probe was used in a colonoscopy.Upon argon plasma coagulation, a perforation occurred in the cecum.Therefore, surgery was performed to repair the perforation.
|