DORNOCH DUO FLUID CART WITH POWER IV; APPARATUS, SUCTION, WARD USE, PORTABLE, AC-POWERED
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Electrical /Electronic Property Problem (1198)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/06/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the unit had power issues and a power inlet module failure.No reported charring, smoke, etc.The event timing was outside of surgery.There was no patient involvement.During device evaluation, it was discovered that the power inlet module was burnt.This investigation discovery led to the final assessment being altered to a reportable malfunction.Due diligence is complete.No additional information is available.
|
|
Manufacturer Narrative
|
This event is recorded with zimmer biomet under (b)(4).This medwatch is being filed as an initial / final report based on information discovered during the device evaluation.Review of the most recent repair record determined the cart was having power issues; the power inlet module was burnt up.The power inlet module was replaced and resolved the reported issue.Review of the device history record identified no deviations or anomalies during manufacturing related to the reported event.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|