• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DENTSPLY LLC AUTOMATE TOOL REFILL; INSTRUMENTS, DENTAL HAND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DENTSPLY LLC AUTOMATE TOOL REFILL; INSTRUMENTS, DENTAL HAND Back to Search Results
Catalog Number 663100
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/20/2023
Event Type  malfunction  
Manufacturer Narrative
There has been a previous report received where this malfunction resulted in a serious injury.Therefore, it must be presumed that recurrence of this malfunction could possibly cause or contribute to a serious injury or require medical or surgical intervention to preclude such.As such, this event is reportable per 21cfr part 803.The device is available for evaluation, though has not been returned as of this report.Evaluation results will be submitted as they become available.
 
Event Description
In this event it is reported that a automate tool refill broke during use.Reportedly, no injury.
 
Manufacturer Narrative
Adding udi # : udi # (b)(4).Returned product was 1 automate iii handle and 1 flexshaft date code 1021 with coil deformation/weld failure causing the product to not function properly.Capa-2021-375 opened to address weld failures for product manufactured by (b)(4) since december 2020 (date code 1220) through implementation date of august 2022 (0822).Complaint is considered substantiated.(nwv).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOMATE TOOL REFILL
Type of Device
INSTRUMENTS, DENTAL HAND
Manufacturer (Section D)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer (Section G)
DENTSPLY LLC
38 west clarke avenue
milford DE 19963
Manufacturer Contact
hannah seevaratnam
221 west philadelphia st.
york, PA 17401
7178457511
MDR Report Key16599461
MDR Text Key312224262
Report Number2515379-2023-00066
Device Sequence Number1
Product Code DZN
UDI-Device IdentifierD0026631001
UDI-PublicD0026631001
Combination Product (y/n)N
Reporter Country CodeNZ
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional
Reporter Occupation Dentist
Type of Report Initial,Followup
Report Date 07/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue Number663100
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Distributor Facility Aware Date03/20/2023
Initial Date Manufacturer Received 03/20/2023
Initial Date FDA Received03/23/2023
Supplement Dates Manufacturer Received03/20/2023
Supplement Dates FDA Received07/26/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Type of Device Usage A
Patient Sequence Number1
-
-