The hillrom technician found the power cord needed to be replaced per the hillrom service manual, perform annual preventive maintenance procedures to make sure all versacare® bed components are functioning as originally designed.Examine the plug for damage.Make sure the plug is a one-piece molded plug assembly.If it is not, replace the plug cord assembly.Replace any plug cord assembly that shows any of these: discoloration of the plug molding, around the plug blades, any signs of cracking., loose fit of the plug blade (the plug blade moves in the molding), verify that the strain relief p-clip is present.Replace the power cord, if damaged.A search of the hillrom maintenance records showed hillrom performed preventative maintenance on this bed on feb 16, 2023.The technician replaced the power cord to resolve the reported event.Based on this information, no further action is required.
|