Model Number 24653
Device Problem Activation, Positioning or Separation Problem (2906)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/02/2023
Event Type  malfunction  
Event Description
It was reported that the handle was disassembled to fully deploy the stent, which constitutes as a partial deployment.An eluvia drug-eluting vascular stent system was selected for use in the endovascular therapy procedure to treat the patient condition of peripheral artery disease.The target lesion was located within the superficial femoral artery and was reported to be 99% stenosed with mild tortuosity and moderate calcification.The system was advanced to the target lesion and stent deployment was attempted.The stent was unable to be deployed using the thumbwheel, but the pull grip was not attempted.The handle was disassembled using a sterile instrument and the stent was manually deployed.The procedure was completed and there were no reported adverse consequences to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELUVIA DRUG-ELUTING VASCULAR STENT SYSTEM
Type of Device
STENT, SUPERFICIAL FEMORAL ARTERY, DRUG-ELUTING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
jay johnson
4100 hamline ave n
arden hills, MN 55112
6515810888
MDR Report Key16600582
MDR Text Key312190564
Report Number2124215-2023-13554
Device Sequence Number1
Product Code NIU
Combination Product (y/n)Y
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/08/2024
Device Model Number24653
Device Catalogue Number24653
Device Lot Number0028941803
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/02/2023
Initial Date FDA Received03/23/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured02/08/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-