MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE); PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS, INC. DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE); PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE) Back to Search Results
Model Number VPR-GW-FLEX14
Device Problem Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 02/23/2023
Event Type  Injury  
Manufacturer Narrative
The guide wire was returned to csi.An extra spring tip was observed on the core shaft.Visual examination did not observe any fracture or damage on the guide wire.The extra spring tip was due to an outside supplier component issue.The reported event of guide wire fracture was not confirmed.The material inspection report for this guidewire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.Csi id: (b)(4).
 
Event Description
Following treatment of a severely calcified lesion, angiographic imaging revealed the tip of the viperwire advance guide wire fractured in the ankle in the posterior tibial artery.A non-csi support catheter was used to retrieve the fractured tip.A shepherd guide wire was used to perform balloon angioplasty.The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 PERIPHERAL ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE GUIDE WIRE)
Type of Device
PERIPHERAL ATHERECTOMY DEVICE (GUIDE WIRE)
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS, INC.
1225 old hwy 8 nw
st. paul MN 55112
Manufacturer Contact
tonia moskalets
1225 old hwy 8 nw
st. paul, MN 55112
MDR Report Key16600902
MDR Text Key311919857
Report Number3004742232-2023-00076
Device Sequence Number1
Product Code MCW
UDI-Device Identifier10852528005718
UDI-Public(01)10852528005718(17)241130(10)463246-1
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190634
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/23/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVPR-GW-FLEX14
Device Catalogue Number7-10041-03
Device Lot Number463246-1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/07/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/23/2023
Initial Date FDA Received03/23/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-