MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ASSURITY MRI; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2272
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The reported event of an incorrect average rate was confirmed.Based on the information provided and analysis, the patient had three episodes exceeding the maximum duration limit of recorded episodes, resulting in an incorrect average rate calculation as the counter does not have a limit.The device is performing per design and no anomalies were found.
 
Event Description
It was reported that the patient presented remotely via merlin.Net.Upon review of transmission, it was found that the pacemaker exhibited incorrect measurements on average heart rate.No interventions were performed.Patient condition was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASSURITY MRI
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16608141
MDR Text Key312021586
Report Number2017865-2023-15629
Device Sequence Number1
Product Code LWP
UDI-Device Identifier05414734509589
UDI-Public05414734509589
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 03/24/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2019
Device Model NumberPM2272
Device Catalogue NumberPM2272
Device Lot NumberP000058779
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/20/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-