ARROW INTERNATIONAL LLC ARROW HEMODIALYSIS KIT: 2-L 15 FR X 27 CM; CATHETER, HEMODIALYSIS, IMPLA
|
Back to Search Results |
|
Model Number IPN920484 |
Device Problem
Obstruction of Flow (2423)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Qn#(b)(4).
|
|
Event Description
|
It was reported: "after the device placement under fluoroscopy, while testing the two ends of the catheter, the one (blue) was working perfect while at the other tip (red), one could with extremely high pressure only insert saline." customer reported only the red extension line was not working properly after placement.The device was removed and "there is a big amount of clot blocking the one of the two end holes".Another device was placed.The patient's condition is reported as fine.
|
|
Event Description
|
It was reported: "after the device placement under fluoroscopy, while testing the two ends of the catheter, the one (blue) was working perfect while at the other tip (red), one could with extremely high pressure only insert saline." customer reported only the red extension line was not working properly after placement.The device was removed and "there is a big amount of clot blocking the one of the two end holes".Another device was placed.The patient's condition is reported as fine.
|
|
Manufacturer Narrative
|
(b)(4).The customer returned multiple components from a hemodialysis kit including a connector assembly, catheter body, two dilators, and sheath.Visual inspection of the connector assembly and catheter body did not reveal any defects or anomalies.The outer diameter of the arterial line measured 0.1880", which is within specifications of 0.185-0.191" per extrusion graphic.The inner diameter of the arterial line measured 0.123", which is within specifications of 0.120-0.126" per extrusion graphic.The connector assembly was functionally tested per the instructions for use (ifu) provided with this kit which states, "flush hub connection assembly with sterile normal saline for injection and clamp extension lines to contain saline within lumen(s)." both lines of the connector assembly were flushed with a lab inventory water filled syringe with not leaks or blocks detected.The ifu also states, "flush catheter thoroughly with sterile normal saline for injection to remove residual blood, post-treatment and before instilling heparin." the catheter body was assembled to the connector assembly and both lines were flushed again.No leaks or blocks were detected.A device history record review was performed, and no relevant findings were identified.The ifu provided with this kit instructs the user, "open catheter clamp prior to infusion through lumen." the complaint of an extension line blocked was not able to be confirmed by a complaint investigation on the returned sample.The returned connector assembly and catheter body passed all relevant dimensional and functional testing.No blocks were able to be detected.A device history record review was performed, and no relevant findings were identified.Based on the customer description and the sample returned, no problem was found on the device.Teleflex will continue to monitor and trend on complaints of this nature.
|
|
Search Alerts/Recalls
|
|
|