Block h6: imdrf device code a0401 captures the reportable event of cutting wire broken.Block h10: the returned ultratome xl was analyzed, and a visual evaluation noted that the cutting wire was broken, and it was not returned.Per media analysis on the provided photo, it showed the device was inside the pouch; however, it was not clear to identify any possible problem or damage that contribute to the reported event.The device was observed under magnification, and the cutting wire was broken, the proximal end of the broken wire was blackened.The unit was cut, in order to check the anchor, and then it was identified that the anchor had remnants of the cutting wire, which suggests that the cutting wire was attached to the anchor.No other problems with the device were noted.The reported event of cutting wire break was confirmed.Upon analysis, it was found that the cutting wire was broken, and blackened.Based on the condition of the device, the problem found could have been generated if there was contact between the device and the scope during energization or if the generator exceeded the maximum of voltage during the procedure.Also, energizing the device prior to performing sphincterotomy can compromise the cutting wire's integrity and cause a premature cutting wire fatigue.It is most likely that procedural or anatomical factors encountered during the use of the device could have affected the device performance and its integrity.Based on all gathered information, the most probable root cause of this complaint is adverse event related to procedure.A labeling review was performed and, from the information available, this device was used per the instructions for use (ifu) / product label.
|