• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ACCENT MRI DR RF; IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ACCENT MRI DR RF; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM2224
Device Problems Incorrect Measurement (1383); Wireless Communication Problem (3283)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/17/2023
Event Type  Injury  
Event Description
It was reported that a patient presented to clinic for follow up.Device interrogation revealed date elective replacement indicator (eri) prior to implant date.The implantable cardioverter defibrillator (icd) was also found with telemetry lockout.No changes or intervention were reported.
 
Event Description
Additional information received notes that pacemaker was explanted and replaced.There were no adverse consequences.
 
Manufacturer Narrative
Reported event of false eri and telemetry lockout were confirmed.The device was at elective replacement indicator (eri) level upon receipt.Analysis of device image indicates telemetry lockout occurred in the field was due to excessive rf telemetry usage in a short period of time.This lockout feature disables rf telemetry for a period to prevent battery drain of device.Session records provided show elective replacement indicator (eri) alert had been falsely triggered and end of service (eos) alert was detected.Further analysis performed found no anomaly.Longevity assessment was performed, and device was in the normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCENT MRI DR RF
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16628524
MDR Text Key312206378
Report Number2017865-2023-15956
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P140033
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/15/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/28/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2015
Device Model NumberPM2224
Device Lot Number4346904
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/14/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/14/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-