Model Number G148
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Pacing Problem (1439)
Patient Problem Asystole (4442)
Event Date 08/08/2022
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy defibrillator (crt-d) system exhibited noise on the right ventricular (rv) channel that was being oversensed which resulting in pacing inhibition of greater than two seconds.It was noted that rv pacing impedances are stable.Technical services provided troubleshooting options.At this time, the device and rv lead remains in service.No adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INOGEN X4 CRT-D
Type of Device
IMPLANTABLE DEVICE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
cashel road
clonmel
EI  
Manufacturer Contact
timothy degroot
4100 hamline avenue north
saint paul, MN 55112
6515826168
MDR Report Key16629691
MDR Text Key312164128
Report Number2124215-2023-14500
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00802526534591
UDI-Public00802526534591
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010012/S341
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date05/21/2020
Device Model NumberG148
Device Catalogue NumberG148
Device Lot Number192678
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2023
Initial Date FDA Received03/28/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/31/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age94 YR
Patient SexFemale
-
-