During device evaluation at olympus, it was found the complaint was confirmed and a boroscope found the channel was kinked.Also, evaluation found the up/down knob was leaking, the scope cover had scratches, the bending section cover adhesive was cracked, and angulation was reduced.This event is under investigation.A supplemental report will be submitted upon receiving additional information.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the reported event is unable to be determined.However, the cause of the event is likely due to the stent was caught by the kink.The event can be detected by following the instructions for use (ifu) which state: operation manual chapter 3 preparation and inspection.Olympus will continue to monitor field performance for this device.
|