• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) GALLANT VR; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CDVRA500Q
Device Problem Incorrect Measurement (1383)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/09/2023
Event Type  malfunction  
Event Description
It was reported that the implantable cardioverter defibrillator exhibited a diagnostic anomaly.The clinician stated that the data on the episode tree did not seem accurate.The episode tree was dated with the last transmission date, but showed data since the last electrogram cleared date and not the last transmission date.No intervention was performed.No patient consequences.
 
Manufacturer Narrative
The complaint of ¿the episode tree is showing the "since" date as the last remote transmission instead of the last cleared date¿ was confirmed from the pdfs in the session record.Per device requirement, the date displayed will be the date of the last session, update, or remote session, not the last cleared date.The device is performing per design.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GALLANT VR
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key16635641
MDR Text Key312246248
Report Number2017865-2023-16112
Device Sequence Number1
Product Code NIK
UDI-Device Identifier05415067031990
UDI-Public05415067031990
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 05/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2023
Device Model NumberCDVRA500Q
Device Catalogue NumberCDVRA500Q
Device Lot NumberS000081530
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/08/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
RIGHT VENTRICLE LEAD
Patient SexFemale
-
-