• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US EXPRESSEW III SUTURE PASSER W/O HOOK; SUTURE/NEEDLE PASSER, REUSABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US EXPRESSEW III SUTURE PASSER W/O HOOK; SUTURE/NEEDLE PASSER, REUSABLE Back to Search Results
Model Number 214140
Device Problems Device-Device Incompatibility (2919); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2023
Event Type  malfunction  
Event Description
It was reported by sales rep that during an arthroscopic shoulder procedure on (b)(6) 2023, it was observed that a needle got stuck in an expressew iii suture passer w/o hook device.During in-house engineering evaluation, it was determined that a needle was found stuck in the shaft of the device.Another like device was used to complete the procedure.There were no adverse patient consequences nor surgical delay reported.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by depuy mitek or its employees that the report constitutes an admission that the device, depuy mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Udi: (b)(4).The expiration date is currently unavailable.Reporter is a j&j sales representative.Investigation summary: the product was returned to mitek for evaluation.Mitek then conducted visual inspection of device received.Upon visual inspection revealed wear marks on the device, the needle was found stuck into the shaft of the device, the white plastic flag was missing.Finally, the jaws do not show structural anomalies.Based on the condition of the needle, a functional test cannot be performed.Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.As part of depuy synthes mitek quality process all devices are manufactured, inspected, and released to approved specifications.According with the visual inspection result, this complaint can be confirmed.He possible root cause for the needle stuck can be attributed when deploying the needle with the jaws open which can lead to a jammed needle.However, this cannot be conclusively determined.As per ifu, it is important to inspect the device prior to use to ensure proper mechanical function and do not use if product is damaged.Also, it is necessary to follow the instructions to cleaning and sterilization process and between uses, lubricate moving parts with the water-solubel lubricant.Applying too much force when grasping tissue will impede passage of the needle and suture.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXPRESSEW III SUTURE PASSER W/O HOOK
Type of Device
SUTURE/NEEDLE PASSER, REUSABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key16637086
MDR Text Key312430629
Report Number1221934-2023-01359
Device Sequence Number1
Product Code LXH
UDI-Device Identifier10886705020218
UDI-Public10886705020218
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/29/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number214140
Device Catalogue Number214140
Device Lot Number58402
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/08/2023
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/10/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-