• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WOLVERINE CORONARY CUTTING BALLOON MONORAIL; CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING Back to Search Results
Model Number 3851
Device Problem Entrapment of Device (1212)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/22/2023
Event Type  malfunction  
Event Description
It was reported that a device entrapment occurred.The target lesion was located in the severely calcified artery.A 10mmx2.50mm wolverine coronary cutting balloon was selected for use.During the procedure, it was noted that the device was stuck with a non-boston scientific guidewire.The catheter and the guidewire were removed as a single unit and the procedure was completed with a different device.No patient complications reported.
 
Manufacturer Narrative
Device evaluated by manufacturer: the device was returned for analysis.A visual examination identified that the balloon was tightly wrapped and had not been subjected to positive pressure.Bunching of the balloon material was noted in the distal balloon cone.All blades were securely bonded and no damage to the blades were noted.A visual and tactile examination of the hypotube shaft found the hypotube severely kinked at 13cm distal to the strain relief.Examination of the wire lumen found that the wire lumen was stretched and bunched approximately 4cm proximal from the tip.A guidewire was unable to be loaded as the wire lumen was stretched and bunched.A visual and microscopic examination found no issue with the marker bands.
 
Event Description
It was reported that a device entrapment occurred.The target lesion was located in the severely calcified artery.A 10mmx2.50mm wolverine coronary cutting balloon was selected for use.During the procedure, it was noted that the device was stuck with a non-boston scientific guidewire.The catheter and the guidewire were removed as a single unit and the procedure was completed with a different device.No patient complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WOLVERINE CORONARY CUTTING BALLOON MONORAIL
Type of Device
CATHETER, PERCUTANEOUS TRANSLUMINAL CORONARY ANGIOPLASTY (PTCA), CUTTING/SCORING
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC IRELAND LIMITED
ballybrit business park
galway
EI  
Manufacturer Contact
jeff wallner
4100 hamline ave n
arden hills, MN 55112
6515811560
MDR Report Key16638056
MDR Text Key312244745
Report Number2124215-2023-14505
Device Sequence Number1
Product Code NWX
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/30/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3851
Device Catalogue Number3851
Device Lot Number0031036534
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/22/2023
Initial Date FDA Received03/29/2023
Supplement Dates Manufacturer Received05/19/2023
Supplement Dates FDA Received05/30/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/13/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-