Other text: d10.Device available for evaluation, h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.G1,2 email is: regulatory.Responses@icumed.Com.One device was received.Per visual inspection, the water tank cover was cracked, and drain/quick connector was corroded.Per functional testing, the reservoir was leaking from both side and the bottom of the water tank cover.The complaint was confirmed.The root cause was a worn gasket.It was unknown what caused the condition.The product's history records were reviewed and there were no non-conformances nor service-related issues that would have resulted in the reported complaint.Replaced the reservoir gasket, water tank cover drain/quick connector fitting.Replaced line cord due to it failed electrical test.Perform preventative maintenance (pm) and calibrated.The device passed all functional and delivery tests.
|