Model Number 71992-01 |
Device Problems
Unable to Obtain Readings (1516); Device Displays Incorrect Message (2591)
|
Patient Problems
Hypoglycemia (1912); Discomfort (2330); Cognitive Changes (2551)
|
Event Date 03/27/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.At this time product has not yet been returned.An extended investigation has been performed for the reported complaint and it has been determined that there was no indication that the product did not meet specification.The device history records (dhrs) for the fs libre sensor and fs libre sensor kit were reviewed and the dhrs showed the fs libre sensor and sensor kit passed all tests prior to release.If the product is returned, a physical investigation will be performed and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the abbott diabetes care (adc) device.Customer received a "replace sensor" message and was unable to obtain readings.As a result, customer experienced "general discomfort", "agitation", "cold", and was unable to self-treat, requiring third-party treatment of "sugar with water" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
|
Manufacturer Narrative
|
Sensor (b)(6) has been returned and investigated.The sensor plug is fully seated and no issues were observed.Extracted data from the returned sensor using approved software.The sensor was found to be in sensor state 6 (indicating early termination).Visual inspection has been performed on the sensor plug assembly, no failure modes were observed.Sim vivo testing (simulation of the electrical signal produced by the sensor tail) was performed and all results were within specification.This issue is therefore nonconfirmed.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
An error message was reported with the abbott diabetes care (adc) device.Customer received a "replace sensor" message and was unable to obtain readings.As a result, customer experienced "general discomfort", "agitation", "cold", and was unable to self-treat, requiring third-party treatment of "sugar with water" by a non-healthcare professional.There was no report of death or permanent impairment associated with this event.
|
|
Search Alerts/Recalls
|