• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIRGAS THERAPEUTICS LLC ISPAN SULFUR HEXAFLUORIDE (SF6) GAS; INTRAOCULAR GAS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIRGAS THERAPEUTICS LLC ISPAN SULFUR HEXAFLUORIDE (SF6) GAS; INTRAOCULAR GAS Back to Search Results
Catalog Number 8065797004
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/04/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A physician reported that ophthalmic operating console was not dispensing sulphur hexafluoride gas and gas bubble in patient eye was reduced to unusual size it was noticed on post operative day eight.There were no details of the procedure.There was no report of the patient harm.Additional information has been requested none received till date.
 
Manufacturer Narrative
A check of the batch production record could not be performed because the lot number was not reported.The expiration date could not be determined.A check of the complaint records could not be performed because the lot number was not provided.A check of confirmed complaints for sf6 eye gas not lasting as long in the patient's eye as expected showed no complaints.The sample was not returned.Testing could not be performed.No sample was returned for evaluation.With no additional, related information provided, the reported event was not confirmed.At this time, the root cause of the reported event cannot be determined.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISPAN SULFUR HEXAFLUORIDE (SF6) GAS
Type of Device
INTRAOCULAR GAS
Manufacturer (Section D)
AIRGAS THERAPEUTICS LLC
6141 easton road
p.o. box 310
plumsteadville PA 18949
Manufacturer (Section G)
AIRGAS THERAPEUTICS LLC
6141 easton road
p.o. box 310
plumsteadville PA 18949
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key16643265
MDR Text Key312302763
Report Number2518435-2023-00007
Device Sequence Number1
Product Code LPO
UDI-Device Identifier00380657970049
UDI-Public00380657970049
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P900067
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065797004
Device Lot NumberASKU
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/03/2023
Initial Date FDA Received03/30/2023
Supplement Dates Manufacturer Received06/02/2023
Supplement Dates FDA Received06/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CONSTELLATION VISION SYSTEM.; CONSTELLATION VISION SYSTEM.
-
-