Model Number M00562671
Device Problems Failure to Cut (2587); Defective Device (2588)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/06/2023
Event Type  malfunction  
Event Description
This report pertains to one of five sensation short throws used in the same procedure.It was reported to boston scientific corporation that a sensation medium oval flexible snare was used during an endoscopic mucosal resection procedure performed on (b)(6) 2023.During the procedure, when the handle was tightened all the way, the polyp could not be removed.It was noted that the outer sheath at the front end of the device was found to be shorter when fully withdrawn when compared to the device used to complete the procedure.The procedure was completed with a similar sensation short throw.There were no patient complications reported as a result of this event.
 
Manufacturer Narrative
E1: initial reporter facility: (b)(6) hospital.E1: initial reporter city: (b)(6).H6: imdrf device code a050702 captures the reportable event of loop unable to cut.
 
Manufacturer Narrative
Block e1: (initial reporter facility) (b)(6) hospital.Block e1: (initial reporter city) (b)(6).Block h6: imdrf device code a050702 captures the reportable event of loop unable to cut.
 
Event Description
Note: this report pertains to one of five sensation short throws used in the same procedure.It was reported to boston scientific corporation that a sensation medium oval flexible snare was used during an endoscopic mucosal resection procedure performed on (b)(6) 2023.During the procedure, when the handle was tightened all the way, the polyp could not be removed.It was noted that the outer sheath at the front end of the device was found to be shorter when fully withdrawn when compared to the device used to complete the procedure.The procedure was completed with a similar sensation short throw.There were no patient complications reported as a result of this event.
 
Manufacturer Narrative
Block e1: (initial reporter facility) (b)(6).Block e1: (initial reporter city) (b)(6).Block h6: imdrf device code a050702 captures the reportable event of loop unable to cut.Block h10 investigation results: one sensation snare was received for analysis.Upon visual inspection, there were no damages found.During functional evaluation, the device showed that it could extend and retract well when connected to the 10-inch loop fixture.During the dimensional test, the device was within specification.No other issues were observed.The reported event of "loop failure to cut" could not be confirmed, as it was not possible to evaluate the device's function during the specific procedure.It is possible that procedural factors, such as physician technique or handling of the device during initial use or setup, may have contributed to the device's condition.The product record review revealed that this was not a new failure type, and the risk had been anticipated.There was no evidence of a manufacturing, design, or user problem that could have caused the complaint.Device analysis showed no issues during visual, functional, and dimensional testing.Based on the available information and device analysis, the most probable root cause for the reported complaint is "no problem detected." a review of the device history record (dhr) confirmed that the device met all material, assembly, and performance specifications at the time of release for distribution.
 
Event Description
Note: this report pertains to one of five sensation short throws used in the same procedure.It was reported to boston scientific corporation that a sensation medium oval flexible snare was used during an endoscopic mucosal resection procedure performed on (b)(6) 2023.During the procedure, when the handle was tightened all the way, the polyp could not be removed.It was noted that the outer sheath at the front end of the device was found to be shorter when fully withdrawn when compared to the device used to complete the procedure.The procedure was completed with a similar sensation short throw.There were no patient complications reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSATION SHORT THROW
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key16645499
MDR Text Key312512786
Report Number3005099803-2023-01393
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729158141
UDI-Public08714729158141
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K131700
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/30/2024
Device Model NumberM00562671
Device Catalogue Number6267
Device Lot Number0027239139
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/07/2023
Initial Date FDA Received03/30/2023
Supplement Dates Manufacturer Received04/05/2023
06/05/2023
Supplement Dates FDA Received04/24/2023
06/27/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/01/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age56 YR
Patient SexMale
Patient Weight74 KG
-
-