The mayfield skull clamp (a1059) was returned for evaluation: failure analysis - the reported complaint was confirmed from the evaluation.The skull clamp had rotational movement in its swivel lock assembly and needed to be cleaned, and some worn off small parts must be replaced.To resolve the issues, the required small parts were replaced and the swivel lock assembly was overhauled.After reassembling of the unit, a successful function check was performed.Root cause -the swivel lock assembly had rotational movement and required replacement of worn components due to routine use and wear.Additionally, improper or suboptimal placement of the skull clamp can contribute to slippage or movement of the patient¿s head.No corrective action has been initiated for the same or similar issue.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
|