It was reported that the patient presented to the hospital for an implantable cardioverter defibrillator (icd) implant procedure on (b)(6) 2023.While attempting to implant the icd, it was noted that there was a faulty setscrew on the device.After multiple failed attempts, the anomalous icd was removed and replaced without further incident in the same procedure.The patient was in stable condition.
|
The reported field event of set screw anomaly was confirmed.The right ventricular set screw contained septum material inside the hex cavity.This material prevented full insertion of the torque driver and was the cause of the reported event.The set screw anomaly was consistent with having occurred during the procedure.Telemetry, impedance, sensing, pacing and high voltage (hv) output functions of the device were tested and found to be normal.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
|