• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL AMPLATZER TALISMAN PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL AMPLATZER TALISMAN PFO OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-PFO-2518
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Atrial Fibrillation (1729)
Event Date 01/14/2023
Event Type  Injury  
Manufacturer Narrative
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
Clinical information: 806- clinical name, patient site id: (b)(6).It was reported that on (b)(6) 2023, a 18x25mm amplatzer talisman pfo occluder was implanted into a patient using an 8f trevisio delivery system.On (b)(6) 2023, the patient had a loop recorder implanted.The transmissions were reviewed and a 1.5 minute atrial fibrillation episode was discovered.Medication was adjusted.The patient's status at the time of report was noted as stable.It was noted that the patient had a history atrial fibrillation in the past.No additional information was provided.
 
Manufacturer Narrative
An event of atrial fibrillation was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.Based on the information received, the cause of the reported incident could not be conclusively determined.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.There is no indication of a product quality issue with respect to labeling design or manufacturing of the device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER TALISMAN PFO OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
ABBOTT MEDICAL
5050 nathan lane n
plymouth MN 55442
Manufacturer (Section G)
ABBOTT MEDICAL REG# 2135147
5050 nathan ln n
plymouth MN 55442
Manufacturer Contact
karen krouse
5050 nathan lane n
plymouth, MN 55442
6517565400
MDR Report Key16653367
MDR Text Key312406784
Report Number2135147-2023-01408
Device Sequence Number1
Product Code MLV
UDI-Device Identifier05415067033314
UDI-Public05415067033314
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120021
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 06/09/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2023
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9-PFO-2518
Device Catalogue Number9-PFO-2518
Device Lot Number8230702
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received06/08/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/19/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age55 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-