• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Model Number A1059
Device Problem Device Slipped (1584)
Patient Problems Laceration(s) (1946); Unspecified Tissue Injury (4559)
Event Date 02/28/2023
Event Type  Injury  
Manufacturer Narrative
An investigation has been initiated based on the reported information.Upon completion of the investigation, a follow-up report will be submitted.
 
Event Description
A facility reported that the mayfield modified skull clamp (a1059) slipped, and an unspecified injury occurred.Surgical delay is unknown and details of the injury have not been provided.Additional information has been requested.
 
Manufacturer Narrative
Additional information received as follows: the patient sustained a scalp laceration which was stapled with about 2-3 staples, and the patient is doing fine (at least from the cut on head).Investigation findings: the mayfield modified skull clamp (a1059) was returned for evaluation: device history record (dhr) - the dhr was reviewed and shows no abnormalities related to the reported failure.Failure analysis - evaluation found no device deficiencies that would have contributed to the reported complaint.The unit was returned for preventive maintenance (pm).As a result, the rroundhead screw, set screw, label, bolt, nut, washers, o-ring, ball bearings, wave springs and helicoils replaced as part of pm services.Root cause - probable root cause of the reported complaint is improper or suboptimal placement of the skull clamp.No further investigation is required based on the acceptability of risk and no adverse trends were identified.This will be monitored and trended going forward.At present, we consider this complaint to be closed.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH
Manufacturer (Section G)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
cincinnati OH
Manufacturer Contact
vivian nelson
1100 campus drive
princeton, NJ 
6099362319
MDR Report Key16658810
MDR Text Key312462863
Report Number3004608878-2023-00052
Device Sequence Number1
Product Code HBL
UDI-Device Identifier10381780253457
UDI-Public10381780253457
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA1059
Device Catalogue NumberA1059
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 03/06/2023
Initial Date FDA Received03/31/2023
Supplement Dates Manufacturer Received04/27/2023
Supplement Dates FDA Received05/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-