Model Number WA53005A
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/20/2023
Event Type  malfunction  
Event Description
The customer reported to olympus, the telescope, 10 mm, 30°, hd, quick lock had a cloudy lens, and the lower left corner of the screen is hard to see.The issue was found during inspection for use for a therapeutic, laparoscopic colectomy procedure.The procedure was completed with a similar device.Inspection and testing of the returned device found cover glass damage.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
The device was returned to olympus for evaluation.The device evaluation found cover glass damage, image failure, and misalignment of the objective lens.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to user error, improper handling, and application of excessive force.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TELESCOPE, 10 MM, 30°, HD, QUICK LOCK, AUTOCLAVABLE
Type of Device
TELESCOPE, 10 MM, 30°, HD, QUICK LOCK
Manufacturer (Section D)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg, hamburg 22045
GM  22045
Manufacturer (Section G)
OLYMPUS WINTER & IBE GMBH
kuehnstrasse 61
hamburg
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key16659582
MDR Text Key312519217
Report Number9610773-2023-00938
Device Sequence Number1
Product Code EWY
UDI-Device Identifier04042761052764
UDI-Public04042761052764
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K950076
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 07/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberWA53005A
Device Catalogue NumberWA53005A
Device Lot Number507096
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/01/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/01/2023
Initial Date FDA Received03/31/2023
Supplement Dates Manufacturer Received06/21/2023
Supplement Dates FDA Received07/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/08/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-