Model Number WA53005A |
Device Problem
Break (1069)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/20/2023 |
Event Type
malfunction
|
Event Description
|
The customer reported to olympus, the telescope, 10 mm, 30°, hd, quick lock had a cloudy lens, and the lower left corner of the screen is hard to see.The issue was found during inspection for use for a therapeutic, laparoscopic colectomy procedure.The procedure was completed with a similar device.Inspection and testing of the returned device found cover glass damage.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation.The device evaluation found cover glass damage, image failure, and misalignment of the objective lens.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, it is likely that the event occurred due to user error, improper handling, and application of excessive force.Olympus will continue to monitor field performance for this device.
|
|