SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC VENTILATORS PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
|
Back to Search Results |
|
Model Number P310NUS |
Device Problem
No Flow (2991)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
It was reported that the oxygen is not working properly.No patient injury was reported.
|
|
Manufacturer Narrative
|
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Tamper seal is broken.Device was received in with a damaged front panel and missing the end cap for the relief pressure knob.All small knobs were cracked.The oxygen knob is loose on the spindle.Performed oxygen concentration checks using a replacement functioning control knob.The oxygen control knob does not function as intended.Device is scheduled for its routine maintenance.The root cause of the reported issue was found to be customer user.Actions were taken to mitigate the reported issue: replaced the oxygen control knob and installed medical replacement battery (mri) battery, and sintered filter.
|
|