SMITHS MEDICAL INTERNATIONAL, LTD. PNEUPAC VENTILATORS PARAPAC PLUS; VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR) Back to Search Results
Model Number P310NUS
Device Problem No Flow (2991)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the oxygen is not working properly.No patient injury was reported.
 
Manufacturer Narrative
This mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).A product sample was received for evaluation.Visual and functional testing were performed.Tamper seal is broken.Device was received in with a damaged front panel and missing the end cap for the relief pressure knob.All small knobs were cracked.The oxygen knob is loose on the spindle.Performed oxygen concentration checks using a replacement functioning control knob.The oxygen control knob does not function as intended.Device is scheduled for its routine maintenance.The root cause of the reported issue was found to be customer user.Actions were taken to mitigate the reported issue: replaced the oxygen control knob and installed medical replacement battery (mri) battery, and sintered filter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PNEUPAC VENTILATORS PARAPAC PLUS
Type of Device
VENTILATOR, EMERGENCY, POWERED (RESUSCITATOR)
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
bramingham business park, enterprise way
luton, bedforshire LU3 4 BU
UK  LU3 4BU
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16660757
MDR Text Key312507938
Report Number3012307300-2023-03358
Device Sequence Number1
Product Code BTL
UDI-Device Identifier35019315107263
UDI-Public35019315107263
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123957
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 04/01/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/01/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberP310NUS
Device Catalogue NumberP310NUS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/25/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/08/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/24/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-