• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON ENDO-SURGERY, LLC. HARMONIC SCALPEL HAND PIECE; INSTRUMENT, ULTRASONIC SURGICAL Back to Search Results
Catalog Number HP054CN
Device Problem Device Markings/Labelling Problem (2911)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/14/2023
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Batch # a9c041.An analysis of the product could not be performed since a physical sample was not received for evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.However, if the product is received at a later date, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device batch a9c041 and no non-conformances were identified.Photo analysis this is an analysis of a set of images submitted for evaluation.Image 1(hp054cn): the image provided by the customer shows a top view of an hp054 packaging box.Image 2(hp054cn): the image provided by the customer shows a packaging label where the manufacturing date is missing.Instead of the manufacturing date a product code was placed.Image 3(hp054cn): the image provided by the customer shows a lower view of an hp054 packaging box.Based on the photo, the reported event was confirmed.Based on the information currently available, a product issue was identified during the investigation of the sample received.Because the instrument was not returned our evaluation is limited.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
It was reported that there is a label printing error in date of manufacture of box during incoming inspections.The device was not used on patient.There was no patient consequence reported.No additional information can be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HARMONIC SCALPEL HAND PIECE
Type of Device
INSTRUMENT, ULTRASONIC SURGICAL
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*  00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
*   00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
*  
3035526892
MDR Report Key16663968
MDR Text Key312526456
Report Number3005075853-2023-02159
Device Sequence Number1
Product Code LFL
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K002906
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Other
Type of Report Initial
Report Date 04/03/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberHP054CN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/15/2023
Initial Date FDA Received04/03/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/12/2023
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
GENERATOR
-
-