Catalog Number D2C4009K
Device Problem Material Rupture (1546)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that a large volume folfusor burst during filling.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
E1: initial reporter facility name: (b)(6).E1: initial reporter city: (b)(6).The device has been received and the evaluation is in progress.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H10: the actual device was received for evaluation.A visual inspection was performed via the naked eye which observed the bladder had been ruptured.There were no signs of abnormality found on the bladder that may have led to the rupture.The reported condition was verified.The cause of the condition could not be determined, however, a possible cause may be due to inherent variation in the material, as bladders are manufactured with rubber.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOLFUSOR
Type of Device
PUMP, INFUSION, ELASTOMERIC
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - IRVINE
17511 armstrong avenue
building 3
irvine CA 92614
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key16665892
MDR Text Key313338942
Report Number1416980-2023-01490
Device Sequence Number1
Product Code MEB
UDI-Device Identifier00085412474328
UDI-Public(01)00085412474328
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 05/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberD2C4009K
Device Lot Number22M032
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/21/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 03/08/2023
Initial Date FDA Received04/03/2023
Supplement Dates Manufacturer Received05/01/2023
Supplement Dates FDA Received05/05/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/18/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-