It was reported that on (b)(6) 2023, a 27mm navitor valve was chosen for implant.The annulus perimeter was measured to be 77mm and it was noted that the patient had a horizontal aorta.A pre-dilation was performed using a 23mm non-abbott balloon.The valve was partially deployed and re-captured three times due to difficulty positioning the valve in the patient anatomy and instability of the pig-tail catheter.On the third attempt at positioning, the valve was deployed at 1mm and released at 3mm above the annulus.After it was released, the valve migrated above the aortic annulus.The valve was not blocking the coronary arteries but, a lasso (snare) was used to raise the valve further so it would not block the coronary arteries in the future.A second 27mm navitor valve was then implanted and released in an optimal position.A minimal residual leak was present after the valve-in-valve procedure was performed and the coronary arteries were not blocked.The patient remained hemodynamically stable throughout the procedure and there was no clinically significant delay.The patient was discharged to home.No additional information was provided.
|
An event of device migration was reported.A returned device assessment could not be performed as the device remains implanted and was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.Please note per the instructions for use, " once the valve is fully deployed, repositioning and retrieval of the valve is not possible.Attempted retrieval (e.G., use of a guidewire, snare, or forceps) may cause aortic root, coronary artery, and/or myocardial damage." also- " caution: do not re-sheath the valve more than two times.If additional positioning attempts are needed, completely re-sheath the valve and remove the valve from the patient.Use a new valve and delivery system to complete the procedure.".
|