A customer reported that during a patient procedure, using a glidescope core 10-inch monitor, the monitor intermittently indicated that there was no connection to the laryngoscope and would cause the image to go out in the middle of the procedure.No delay during the procedure, use of a backup device, or harm to the patient was reported.
|
The customer declined the option to have their glidescope core 10-inch monitor returned for evaluation.However, upon following up with the customer, they reported that they resolved the issue by inserting the cable into a different port.The customer determined that the cause of the reported connection issue was someone had switched the cable to the wrong port of the monitor.Corrective action is currently not required at this time.Verathon will continue to monitor for any ongoing trends.
|