• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. CADD MEDICATION CASSETTES; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. CADD MEDICATION CASSETTES; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number 21-7346-24
Device Problem Volume Accuracy Problem (1675)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/04/2022
Event Type  malfunction  
Event Description
It was reported that there was quite a bit of medication in the bag after infusion.No patient injury was reported.
 
Manufacturer Narrative
Other, other text: this remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).No product or photographic evidence were provided to aid in this investigation.The complaint report offered insufficient details to determine whether this product functioned as intended or was used in a manner consistent with its instructions for use (ifu) or failed to meet product specifications.Lacking any additional evidence, this complaint has been closed as unconfirmed.If the product is returned, will reopen this complaint for further investigation.
 
Manufacturer Narrative
One device was received inside a plastic bag without its original packaging.No damage or other defects were detected during visual inspection.No discrepancies were detected during functional testing and successfully passed with a result within spec -1.74% (average delivery error acceptable is -7% to 7%).The complaint was not confirmed.A device history record (dhr) review reported no discrepancies or non-conformances during the manufacturing of the reported lot number.No action taken since complaint was not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CADD MEDICATION CASSETTES
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16676183
MDR Text Key313399608
Report Number3012307300-2023-03504
Device Sequence Number1
Product Code FPA
UDI-Device Identifier15019517161953
UDI-Public15019517161953
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 10/13/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/04/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number21-7346-24
Device Catalogue Number21-7346-24
Device Lot Number4089487
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received09/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/30/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-