It was reported that the stent did not deploy in the intended location and appeared longer once deployed than on pre-deployment imaging.An innova vascular stent was selected for use during the percutaneous angioplasty procedure.The target location was located within the mid superficial femoral artery (sfa) and was reported to be 75% stenosed with moderate-to-severe calcification.The stent delivery system was advanced to the target lesion, and the stent was positioned by identifying both the proximal and distal ends on imaging.The stent was deployed slowly and deliberately, and the proximal end landed far more proximal than the pre-deployment imaging had indicated.The stent appeared much longer than originally visualized.Following stent deployment, a potential arteriovenous fistula (avf) was observed laterally.The avf was not suspected to be related to the innova vascular stent.A charger balloon dilation catheter was inflated to address the avf.The procedure was completed, and there were no reported adverse consequences to the patient.
|