The log file was analysed for investigation.Based on the log file analysis the reported device behaviour could be confirmed.As per log file the device posted an airway pressure negative alarm caused by a performed closed suction.As specified gas delivery stopped and the safety valve opened to ambient to allow spontaneous breathing.However, the log file analysis revealed that the airway pressure negative alarm was not withdrawn after the suction process.The log file entries indicated a deviation in the calibration of the pressure sensors.Therefore, the pba pvt-monitoring (pressure, flow and temperature) and the calibration unit (for the calibration of the pressure sensors) were exchanged and send in for investigation.The pba pvt-monitoring was analysed and tested without any deviation.The examination of the calibration unit revealed a delayed switching behaviour of one valve necessary for calibration the corresponding pressure sensor to ambient pressure.Therefore, the sensor was not correctly calibrated under ambient pressure and measured a pressure less than -1 mbar in normal condition.Since the measured pressure values must be above -1 mbar to cancel the alarm condition, the airway pressure negative alarm was not withdrawn.In the present case the calibration unit already has been replaced.In the present case no patient harm was reported.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|