MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER, 390; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem No Display/Image (1183)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that the warmer display is blank.No report of patient involvement.
 
Manufacturer Narrative
Date of event is unknown, no information has been provided to date.A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other, other text: one device was returned for investigation with a corroded drain fitting, faulty line cord, leaking block assembly, and cracked tank cover.Visual inspection was performed then filled the tank with water, attached temp check, plugged in line cord, and turned on the power switch.Functional tests reported that the display worked but had liquid crystal leakage inside that aloud only part of the lcd to be read.Confirming the reported complaint.A faulty pcb from impact was the cause of the reported issue.After replacing the pcb, the device was able to function as intended.A manufacturing dhr review was not performed because the device is beyond a year from its manufacture date and there was no indication of a manufacturing defect during the investigation.Service history review identified this device has not been in for service previously.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER, 390
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16684075
MDR Text Key312787384
Report Number3012307300-2023-03582
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public10695085002796
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberHL-390
Was Device Available for Evaluation? No
Date Returned to Manufacturer04/03/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/16/2023
Initial Date FDA Received04/05/2023
Supplement Dates Manufacturer Received05/02/2023
Supplement Dates FDA Received05/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-