• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OAKDALE LEVEL 1 HOTLINE FLUID WARMER; WARMER, THERMAL, INFUSION FLUID Back to Search Results
Model Number HL-390
Device Problem No Display/Image (1183)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that the warmer display is blank.No report of patient involvement.
 
Manufacturer Narrative
Date of event is unknown, no information has been provided to date.A product sample was received and is awaiting evaluation, investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.
 
Manufacturer Narrative
Other, other text: h3.Device evaluated by manufacturer and h6.Evaluation codes: updated.One device was returned for investigation.It was noted that the entire device has multiple scuff marks, and the liquid crystal display (lcd) had liquid crystal leakage and the water tank leaked when filled with water, chassis was missing a screw and corroded quick drain connect.Put water in tank, attached temperature check, plugged line cord into outlet, turned device on, brought circulating water up to 42.0 degrees.Functional testing confirmed the customer complaint.A root cause was determined to be due to the device being dropped or sharply impacted.A manufacturing device history review was not performed as the device is a year past the manufacturing date.Service history review identified this device has not been in for service previously.Replaced the printed circuit board (pcb), front cover, water tank cover, quick connected and performed preventative maintenance (pm).Device passed all functional tests.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL 1 HOTLINE FLUID WARMER
Type of Device
WARMER, THERMAL, INFUSION FLUID
Manufacturer (Section D)
OAKDALE
3350 granada ave n
oakdale MN 55128
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16685631
MDR Text Key312733182
Report Number3012307300-2023-03576
Device Sequence Number1
Product Code LGZ
UDI-Device Identifier10695085002796
UDI-Public10695085002796
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K911383
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Distributor
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 05/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberHL-390
Device Catalogue NumberHL-390
Was Device Available for Evaluation? No
Date Returned to Manufacturer04/03/2023
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/16/2023
Initial Date FDA Received04/05/2023
Supplement Dates Manufacturer Received05/02/2023
Supplement Dates FDA Received05/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/07/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-