• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME HANDPIECE, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER SURGICAL, INC. ZIMMER AIR DERMATOME; DERMATOME HANDPIECE, PNEUMATIC Back to Search Results
Model Number N/A
Device Problems Calibration Problem (2890); Power Problem (3010)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported that the device rpms were in specification but erratic.The control bar was in the correct position.The calibration was out at the 0 and 20 reading.The event timing was during pm.There is no harm or delay reported.No adverse events were reported as a result of this malfunction.Due diligence is in progress.No further information is available at this time.
 
Manufacturer Narrative
This complaint is recorded by zimmer biomet under cmp-(b)(4).A follow up/ final report will be submitted once investigation is complete.If further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor the trends.
 
Manufacturer Narrative
If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no additional info reported.
 
Manufacturer Narrative
G2: foreign country- canada/ review of the most recent repair record determined the rpms were in specification but erratic, the control bar was in the correct position, and the calibration was out at the 0 and 20 reading.The neck, housing, swivel, lever, reciprocating arm, motor, eccentric shaft, fine adjustment cams, bearings, gears, spring seal and other parts were replaced and resolved the reported issue.Dhr was reviewed and no discrepancies related to the reported event were found.A definitive root cause cannot be determined.The event is confirmed.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
There is no additional event information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIMMER AIR DERMATOME
Type of Device
DERMATOME HANDPIECE, PNEUMATIC
Manufacturer (Section D)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer (Section G)
ZIMMER SURGICAL, INC.
200 west ohio avenue
dover OH 44622
Manufacturer Contact
jennifer rapsavage
56 e. bell drive
warsaw, IN 46582
5745260384
MDR Report Key16689860
MDR Text Key312803118
Report Number0001526350-2023-00339
Device Sequence Number1
Product Code GFD
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 07/20/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00880100000
Device Lot Number31609500
Was Device Available for Evaluation? Device Returned to Manufacturer
Was the Report Sent to FDA? No
Date Manufacturer Received07/19/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/23/1998
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberN/A
Patient Sequence Number1
Patient SexPrefer Not To Disclose
-
-