MAUDE Adverse Event Report: SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL INTERNATIONAL, LTD. PORTEX GENERAL ANESTHESIA CIRCUITS; CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE Back to Search Results
Catalog Number C37101329-NLJ
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/01/2021
Event Type  malfunction  
Event Description
It was reported that during a pre-use check, leakage of air was observed.No patient injury.No additional information is available for this complaint.
 
Manufacturer Narrative
Udi number and g5 are unknown.Event problem and evaluation codes: updated.A device history record (dhr) review was conducted which indicated the lot was manufactured with 1,650 pieces and was completed with a sort instruction for bad assembly on the tube to the wye.Issue was identified by the operator on the process, nevertheless, this issue is not related with the failure reported by the customer.Three product samples and photos were received for evaluation.Visual inspection and functional testing found only one of the three samples provided it is confirmed as leakage failure for a hole in the breathing circuit due embedded resin on the rings of the circuit.A non-conforming event (nce) was opened along with an internal ncr (b)(6) 2021 due the occurrence of this failure report, and investigation to identify the root cause will be performed.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# 617147.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX GENERAL ANESTHESIA CIRCUITS
Type of Device
CIRCUIT, BREATHING (W CONNECTOR, ADAPTOR, Y PIECE
Manufacturer (Section D)
SMITHS MEDICAL INTERNATIONAL, LTD.
boundary road
hythe, kent CT21 6JL
UK  CT21 6JL
Manufacturer (Section G)
NULL
Manufacturer Contact
jim vegel
MDR Report Key16691493
MDR Text Key313420342
Report Number3012307300-2023-03655
Device Sequence Number1
Product Code CAI
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/09/2023
Device Catalogue NumberC37101329-NLJ
Device Lot Number4051213
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2021
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 04/01/2021
Initial Date FDA Received04/06/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient RaceAsian
-
-