H6: event problem and evaluation codes: updated.H3: device evaluated by manufacturer: updated.H10: a device history record (dhr) review was conducted which indicated all inspections were completed and no issues were noted during manufacture.One sample was returned for investigation.Sample was received in used conditions without its original packaging, decontaminated and inside in a plastic bag.Visual inspection and functional test were performed.Visual inspection found no damage, kinks or any discrepancies that could cause the failure mode reported.For functional testing the sample was filled with 100 ml of water and connected to the cadd legacy plus to look for unusual function.No alarms were activated.The complaint was not confirmed.No root cause was determined as the complaint was not confirmed.No action was taken as the complaint was not confirmed.No fault was found with the device.This remediation mdr was generated under protocol (b)(4), as a result of warning letter cms# (b)(4).
|