The device was returned to olympus for evaluation and repair.The investigation is still in progress; therefore, the root cause of the reported defect cannot be determined at this time.However, if additional information becomes available, this report will be supplemented accordingly.In general, the customer is required to inspect the device for defects, check the function of all devices and have alternate equipment prior to use.Olympus will continue to monitor the field performance of this device.Additional pma/510(k): k931995.
|
This report is being supplemented to provide additional information based on the device evaluation and the legal manufacturer's final investigation.Correction to g2, adding other and china.Information was inadvertently not included on the initial medwatch.The device was returned and an evaluation was completed for it.During inspection, olympus confirmed the reported event, the ceramic tip of the inner sheath was missing.Additionally, the sealing ring was damaged; however, this defect is not considered severe enough to cause a potential adverse event.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 5 years since the subject device was manufactured.Based on the results of the investigation, it is likely the ceramic tip of the inner sheath being missing occurred due to wear and tear and/or improper handling by the customer.However, a definitive root cause could not be determined.The following information is stated in the instructions for use (ifu): ¿4 before use: warning: infection control risk: properly reprocess the product before first and each subsequent use following the instructions in this manual and in the system guide endoscopy.Improper and/or incomplete reprocessing can cause infection of the patient and/or medical personnel.4.1 inspection and testing: inspecting the product: visually inspect the product.Make sure that it has: no corrosion, no dents, and no scratches.Ceramic insulation at distal end: visually inspect the ceramic insulation at the sheath¿s distal end before each use.Do not use the instrument in case of damage (e.G.Cracks, fractures).Warning: risk of injury: impact, fall, shock or similar stress can damage the ceramic insulation at the sheath¿s distal end.Damaged instruments can cause injuries to the patient and/or user.Do not use the instrument if damaged.Damaged product: if the product is damaged or does not function properly, contact an olympus representative or an authorized service center.¿ olympus will continue to monitor field performance for this device.
|