Related manufacturer reference number: 2017865-2023-16924.It was reported the patient presented to the hospital decompensating.Upon interrogation, it was noted the right ventricular lead exhibited a drop in pacing impedance.An echocardiogram showed the right ventricular lead had perforated and a pericardial effusion which required pericardiocentesis.While explanting the right ventricular lead, the helix appeared to be bent.The right ventricular lead was explanted and replaced.When connecting the right ventricular lead to the chronic implantable cardioverter defibrillator (icd), something was blocking the wrench from engaging with the set screw to tighten down the lead.The icd was explanted and replaced.The patient was in stable condition after the procedure.
|