• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 36/+7,5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH DELTA V-40 CERAMIC HEAD 36/+7,5; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS Back to Search Results
Model Number 6570-0-736
Device Problem Corroded (1131)
Patient Problems Memory Loss/Impairment (1958); Pain (1994); Metal Related Pathology (4530)
Event Date 06/06/2019
Event Type  Injury  
Event Description
Rep reported the patient's left hip implanted on (b)(6) 2011 was revised on (b)(6) 2023.Initial patient complaints were pain, elevated ion levels, and memory loss.Intra-operatively, corrosion was noted at the stem/ neck interface and at the interface of the trunnion and ceramic head.The stem construct, head, and liner were revised.
 
Manufacturer Narrative
Reported event: an event regarding corrosion involving a ceramic head was reported.The event was not confirmed.Method & results: product evaluation and results: material analysis, visual, functional and dimensional inspections could not be performed as the device was not returned.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the event could not be confirmed as insufficient information was provided.Further information such as device return, pre- and post-operative x-rays, primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information become available to indicate further evaluation is warranted, this record will be reopened.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DELTA V-40 CERAMIC HEAD 36/+7,5
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/CERAMIC/POLYMER, CEMENTED OR NON-POROUS
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
Manufacturer (Section G)
STRYKER ORTHOPAEDICS-CORK
ida industrial estate
carrigtwohill NA
EI   NA
Manufacturer Contact
arokiya raj
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16697182
MDR Text Key312860566
Report Number0002249697-2023-00364
Device Sequence Number1
Product Code LZO
UDI-Device Identifier04546540608581
UDI-Public04546540608581
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K052718
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 04/06/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number6570-0-736
Device Catalogue Number6570-0-736
Device Lot Number35423602
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 03/15/2023
Initial Date FDA Received04/06/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/18/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age72 YR
Patient SexMale
Patient Weight91 KG
-
-