• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER TRAUMA KIEL UNKNOWN T2 TIBIA NAIL; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER TRAUMA KIEL UNKNOWN T2 TIBIA NAIL; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Catalog Number UNK_KIE
Device Problem Difficult to Remove (1528)
Patient Problem Insufficient Information (4580)
Event Date 03/10/2023
Event Type  malfunction  
Manufacturer Narrative
Once the investigation has been completed any additional information will be reported in a supplemental report.
 
Event Description
As reported by the sales rep, "surgeon encountered difficulties in the removal of a t2 tibia nail.It proved impossible to remove the obturator plug and, by extension, the nail despite the use of stryker extraction kits.A left-handed screwdriver, reference (b)(4), broke off in the plug.".
 
Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling was not possible because the catalog number and lot number were not communicated.Indications of material, manufacturing, or design related problems were unable to be identified as the catalog number and lot number were not communicated.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.H3 other text : device disposition unknown.
 
Event Description
As reported by the sales rep, "surgeon encountered difficulties in the removal of a t2 tibia nail.It proved impossible to remove the obturator plug and, by extension, the nail despite the use of stryker extraction kits.A left-handed screwdriver, reference (b)(4), broke off in the plug.".
 
Manufacturer Narrative
The reported event could not be confirmed since the device was not returned for evaluation and no other evidences were provided.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling was not possible because the catalog number and lot number were not communicated.Indications of material, manufacturing, or design related problems were unable to be identified as the catalog number and lot number were not communicated.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If the device is returned or if any additional information is provided, the investigation will be reassessed.Device disposition unknown.
 
Event Description
As reported by the sales rep, "surgeon encountered difficulties in the removal of a t2 tibia nail.It proved impossible to remove the obturator plug and, by extension, the nail despite the use of stryker extraction kits.A left-handed screwdriver, reference (b)(4), broke off in the plug.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN T2 TIBIA NAIL
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM  D-24232
Manufacturer (Section G)
STRYKER TRAUMA KIEL
prof. kuentscher-strasse 1-5
schoenkirchen/kiel D-242 32
GM   D-24232
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key16697306
MDR Text Key313294480
Report Number0009610622-2023-00107
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 06/07/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/06/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK_KIE
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/10/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-