• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MICROPORT CRM S.R.L. PLATINIUM; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, CARDIAC RESYNCHRONIZATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MICROPORT CRM S.R.L. PLATINIUM; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, CARDIAC RESYNCHRONIZATION Back to Search Results
Model Number PLATINIUM SONR CRT-D 1811
Device Problem Battery Problem (2885)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/23/2023
Event Type  malfunction  
Manufacturer Narrative
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.Analysis of the provided data revealed : a normal battery curve - the device was manufacturer in 2020 - the last battery reforming occurred the 23rd february 2023.The battery voltage is still below 3v several weeks later (on 15 march 2023) - according to the manual, the recommendation is to not implant this device.
 
Event Description
Reportedly, the icd displayed a battery voltage <3v before implantation.The device was not implanted.
 
Manufacturer Narrative
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.A review of manufacturing documents confirmed that the device was manufactured according to specifications analysis of the provided data revealed : - a normal battery curve - the device was manufactured in 2020 - the last battery reforming occurred the 23rd february 2023.The battery voltage is still below 3v several weeks later (on 15 march 2023) - according to the manual, the recommendation is to not implant this device.
 
Event Description
Reportedly, the icd displayed a battery voltage <3v before implantation.The device was not implanted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PLATINIUM
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, CARDIAC RESYNCHRONIZATION
Manufacturer (Section D)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT  13040
Manufacturer (Section G)
MICROPORT CRM S.R.L.
via crescentino s.n
saluggia (vc) 13040
IT   13040
Manufacturer Contact
elodie vincent
via crescentino s.n
saluggia (vc) 13040
IT   13040
146013665
MDR Report Key16700322
MDR Text Key312967648
Report Number1000165971-2023-00259
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 06/05/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPLATINIUM SONR CRT-D 1811
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 03/10/2023
Initial Date FDA Received04/07/2023
Supplement Dates Manufacturer Received05/22/2023
Supplement Dates FDA Received06/05/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient EthnicityNon Hispanic
-
-