Model Number PLATINIUM SONR CRT-D 1811 |
Device Problem
Battery Problem (2885)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/23/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.Analysis of the provided data revealed : a normal battery curve - the device was manufacturer in 2020 - the last battery reforming occurred the 23rd february 2023.The battery voltage is still below 3v several weeks later (on 15 march 2023) - according to the manual, the recommendation is to not implant this device.
|
|
Event Description
|
Reportedly, the icd displayed a battery voltage <3v before implantation.The device was not implanted.
|
|
Manufacturer Narrative
|
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by microport crm that was cleared or approved by fda for marketing in the united states.A review of manufacturing documents confirmed that the device was manufactured according to specifications analysis of the provided data revealed : - a normal battery curve - the device was manufactured in 2020 - the last battery reforming occurred the 23rd february 2023.The battery voltage is still below 3v several weeks later (on 15 march 2023) - according to the manual, the recommendation is to not implant this device.
|
|
Event Description
|
Reportedly, the icd displayed a battery voltage <3v before implantation.The device was not implanted.
|
|