H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of the cable is exposed wires are showing was confirmed.The reported issue is confirmed; the probe¿s cable sleeve is pulling away from the strain relief.The probe¿s cable sleeve is pulling away from the strain relief due to an adhesive failure.H3 other text : evaluation findings are in section h11.
|