DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID, TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Component Incompatible (1108)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 03/29/2023
Event Type  malfunction  
Event Description
It was reported that during preventive maintenance (pm) performed by a getinge service territory manager (stm) , the cardiosave intra-aortic balloon pump (iabp) unit failed fill manifold test.There was no patient involvement.
 
Manufacturer Narrative
A supplemental report will be submitted upon completion of our investigation.
 
Event Description
N/a.
 
Manufacturer Narrative
A getinge field service engineer evaluated during planned maintenance device failed fill manifold test.Replaced helium reservior (997-00-0565), retested device.Device passed with no issues.Verified unit calibrated and passes all functional and safety tests per factory specifications, returned to customer and cleared for clinical use.The defective components were received for further investigation.The failure analysis and testing department was unable to replicate the failure experienced by the customer.Retaining the helium reservoir assembly in the failure analysis and testing department per procedure 0002-07-d008 rev ap.The root cause selection for this investigation will be ¿not confirmed' due to the failure analysis and testing department was unable to replicate the failure.The non-conformances with the returned components were not confirmed.However, the root cause or the most probable root cause is not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID, TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
arelean guzman
1300 macarthur blvd
mahwah, NJ 
MDR Report Key16707102
MDR Text Key312955983
Report Number2249723-2023-01850
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/10/2023
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Manufacturer Received03/12/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/11/2021
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
N/A.
-
-