|
Model Number 24700 |
Device Problem
Difficult to Remove (1528)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 03/14/2023 |
Event Type
Injury
|
Event Description
|
It was reported that removal difficulties requiring surgical intervention occurred.The 75% stenosed target lesion was located in a shunt in a moderately tortuous vessel.A 8.0mm x 40mm x 80cm (4f) sterling balloon catheter was advanced for dilatation.However, during the procedure and inflated two times at 6 atmospheres, the catheter could not be pulled back into the non-boston scientific (bsc) sheath.Inflation and deflation were attempted but resistance was persisted.A small incision was then made, and the device was removed along with the non-bsc sheath.The catheter was intact on the delivery system when the device was removed from the patient no patient complications nor injuries were reported and the patient status was in good condition.
|
|
Manufacturer Narrative
|
Correction to h6: removed impact code for delay to treatment/therapy.Device evaluated by mfr.: returned product consisted of a sterling balloon catheter.The outer shaft, inner shaft, balloon and tip were visually and microscopically examined.Visual examination revealed that the device returned in 3 pieces.The first piece measured approximately 30.5cm from the hub to the separated inflation and guidewire lumen.The inflation lumen of the middle piece is approximately 40cm while the guidewire lumen is approximately 60.6cm long.On the third piece the balloon is separated 40cm from the tip while the guidewire lumen is separated 10cm from the tip.There are multiple kinks along the guidewire lumen.Microscopic examination revealed no additional damages.Inspection of the remainder of the device presented no other damage or irregularities.Product analysis found damage what would have contributed to the difficulty to remove.H6 - impact codes: removed "delay to treatment/therapy".
|
|
Event Description
|
It was reported that removal difficulties requiring surgical intervention occurred.The 75% stenosed target lesion was located in a shunt in a moderately tortuous vessel.A 8.0mm x 40mm x 80cm (4f) sterling balloon catheter was advanced for dilatation.However, during the procedure and inflated two times at 6 atmospheres, the catheter could not be pulled back into the non-boston scientific (bsc) sheath.Inflation and deflation were attempted but resistance was persisted.A small incision was then made, and the device was removed along with the non-bsc sheath.The catheter was intact on the delivery system when the device was removed from the patient no patient complications nor injuries were reported and the patient status was in good condition.
|
|
Search Alerts/Recalls
|
|
|