MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINSONIAN TREMOR
|
Back to Search Results |
|
Model Number B35200 |
Device Problem
Electromagnetic Interference (1194)
|
Patient Problem
Shaking/Tremors (2515)
|
Event Date 04/03/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that the patient's left side got more shaky about the last week or so.Prior to this event, the patient said they got an mri about 10 days or 2 weeks ago and stated the implantable neurostimulator (ins) was turned on during the mri and was not in mri mode because the mri technicians didn't say anything about it.Before calling, the patient increased stimulation for their left side and mentioned that their right side was fine, so they did not adjust the right side.The patient said they will monitor their left side for a couple of days to see if it gets better.
|
|
Event Description
|
Additional information received from the consumer reported the cause of the shaking was them turning stimulation off.The patient was admitted to the hospital for tia symptoms after they had the mri (which resolved itself in 24 hours).To resolve the shaking and potential mri interference the patient couldn¿t have any other mri¿s or turn stimulation off.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|